1. Overall requirements: Class I BF type, with anti-defibrillation protection application part, Class I floating output.
2. Scope of application: Clinically used for pure cutting, mixed cutting, coagulation and bipolar electrocoagulation of soft tissue in surgical operations (except cardiac surgery), and can also be used for electrocuting and electrocoagulation treatment such as endoscopy.
3. Operation mode: intermittent loading continuous operation, operation/interval time is 10s/30s.
4. Control circuit: microprocessor control, full digital circuit display.
5. Host:
5.1 Working power supply: 220V±22V 50Hz±1Hz;
5.2 Working main load frequency: ≤530KHz±2%; Modulation wave frequency: ≤20KHz±5%;
5.3 Power consumption of the whole machine: ≤1400VA;
5.4 Ambient temperature: +10ºC~+40ºC; Relative humidity: 30%~75%; Atmospheric pressure range: 700hpa~1060hpa;
5.5 Storage ambient temperature: -40ºC~55ºC; Storage relative temperature: ≤93%;
★5.6 Service life of the host: ≥8 years;
5.7 Compliance with standards and requirements: The implementation standards of the host meet the national safety standards and medical electrical safety, high-frequency surgical equipment safety, electromagnetic compatibility and other requirements.
6. Main functions and technical parameters:
6.1 The power-on automatic detection system is built in. If the power-on self-detection system is abnormal, all outputs will be automatically cut off immediately to ensure that the host is always in the best state. During the self-detection process, it will automatically detect whether its connected accessories are connected normally;
6.2 The host control panel can be used to control pure cutting, mixed cutting 1, mixed cutting 2, electrocoagulation 1, electrocoagulation 2 and bipolar coagulation modes, and the power output requirements can be adjusted to achieve the best cutting and coagulation effects. Can be connected to various laparoscopes and endoscopes to meet surgical needs;
6.3 Equipped with a patient loop electrode monitoring (REM) system, with single and double-piece neutral electrode automatic identification function; once the electrode fits the patient's skin less than 60%, the system automatically sounds an alarm and stops output at the same time to reduce the risk of negative plate burns and ensure surgical safety;
★6.4 Equipped with a special medical foot switch (with independent registration certificate documents), single and bipolar functions can be controlled by the same foot switch;
6.5 With ≥4 fault code display functions, so as to quickly determine the cause of the fault and ensure surgical safety;
6.6 With automatic adjustment technology, it can control all modes and effects;
6.7 With power off Automatic memory function, which can automatically display the last used mode and power after turning on;
6.8 Equipment protection: long-term operation, with self-protection functions of over-power, over-current and leakage;
6.9 Working modes ≥6; maximum output power ≥300W;
6.10 Rated output power: monopolar pure cutting: ≥300W; monopolar mixed cutting 1: ≥250W; monopolar mixed cutting 2: ≥200W; monopolar electrocoagulation 1: ≥120W; monopolar electrocoagulation 2: ≥100W; bipolar coagulation: ≥70W;
6.11 The output power is step-by-step adjustable, and the minimum adjustable range is: ±1W;
6.12 Output power adjustment mode ≥2, with manual switch and foot switch.
3 monopolar cutting modes: pure cut, blend 1, blend 2
2monopolar coagulation modes: spray coagulation, forced coagulation
1 bipolar output modes: standard coag
CQM contact quality monitoring system
Electrosurgical pens and foot switch control
Start with the recently used mode, power and other parameters
Volume adjustment function;
Cut and coagulate in an intermittent manner.
Properties |
The Basis of Surgical Instruments |
Function |
Cut the tissue and hemostasis |
Operating Frequency |
530 KHz |