| Customization: | Available |
|---|---|
| Classification: | Chemistry Analyzer |
| Type: | Biochemical Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency




Technical Specifications:
Assay Model: Random access with direct reading system
Assay Item: 36 assay
Throughput: 1Reagent:220 test/h(max); 2 Reagent: 160 test/h; With ISE: +100 test/h
Sample Position: 50
Sample Volume: 1-100ul,0.1ul/step
Reagent Position:36
Reagent Volume: R1:1-400ul,1ul/step,R2: 1-400ul,1ul/step
Wavetength: 340-810nm
Light Source: Halogen lamp
Method: End point, Kinetic, Fixed time, Immunoturbidimetry, 1-2 reagen,Multistandard, reagent/serum blank, etc
Quality Control: Levy Jennings quality control program at 3 levels.
Printer: Multiple report formats available
| Model | LHA-6 |
| Operation | Random access |
| Reading System | Direct reading system |
| Method | End point, Kinetic, Fixed time, Bi-chromatic, Immunoturbidimetric, 1-2 reagent multistandard, reagent/serum blank, etc |
| Assay Item | 36 assay items |
| Throughput | 1 Reagent: 220 test/h (max); 2 Reagent: 160 test/h |
| Sample Position | 50 postions , including standard, control stat; either serum cup or primary tube available |
| Sample Volume | 1-100µL, 0.1µL/step |
| Reagent Position | 36 |
| Reagent Volume | R1 : 1µL-400µL, 1µL/step R2 : 1µL-400µL, 1µL/step |
| Reagent/Sample Probe | With liquid level sensor, collision sensor , coating and automatic washing station |
| Reaction Time | 0-999 seconds |
| Cuvette | High quality UV-transmitted plastic cuvette, or quartz glass cuvette on request |
| Cuvette Washing | 5 channels automatic washing system for cuvettes |
| Alarm | Automatic alarm for malfuction (e.g: collision, reagent and distilled water insufficiency, waste container ful, etc) |
| Calibration | Linear, Non-linear, Multi point, K factor |
| Sample Dilution / Retest | Sample can be diluted and retested if results are out of range |
| Wavelength | 340, 405, 450, 505, 546, 578, 620, 670nm(standard), 330, 380, 492, 510, 600, 630, 650, 660, 700, 750, 800, 810nm (optional two filters) |
| Light Source | Halogen lamp |
| Absorbance Resolution | 0.0001A |
| Absorbance Accuracy | 0-0.5A: ±0.02A; 0.5-1.0A: ±0.04A; 1.0-1.5A: ±0.06A |
| Repeatability(CV%) | ALT<3%, UREA<2.5%, TP<1.5% |
| Quality Control | Levy jennings quality control program at 3 levels |
| Bar Code | Optional |
| Printer | Multiple report formats available |
| Other | Tele-diagnostic, bi-direction interface to LIS syste, etc |
| Language | English , French , Spanish , German , Russian , etc |
| System | Window XP or Vista ; Window 7 |
| Power Supply | 220V/50Hz, or 110V/60Hz |
| Dimension | 65 x 53 x 52cm |
| Weight | 45 Kg |